A method to evaluate the performance of X-ray imaging scintillators by means of the brightness-sharpness index (BSI).

ΤίτλοςA method to evaluate the performance of X-ray imaging scintillators by means of the brightness-sharpness index (BSI).
Publication TypeJournal Article
Year of Publication1999
AuthorsCavouras, D., Kandarakis I., Prassopoulos P., Kanellopoulos E., Nomicos C. D., & Panayiotakis G. S.
JournalActa Radiol
Volume40
Issue2
Pagination211-6
Date Published1999 Mar
ISSN0284-1851
Λέξεις κλειδιάHumans, Luminescent Measurements, Radiographic Image Enhancement, Technology, Radiologic, X-Ray Intensifying Screens
Abstract

PURPOSE: To propose an image quality index, the brightness-sharpness index (BSI), for assessing the quality of the image produced by phosphors of medical imaging detectors.
MATERIAL AND METHODS: BSI was evaluated by experimental X-ray luminescence and modulation transfer function measurements. BSI was determined for a number of test phosphor screens prepared from Gd2O2S:Tb, La2O2S:Tb, and Y2O2S:Tb phosphor materials. The screens covered a wide range of coating thicknesses from 50 to 150 mg/cm2 and measurements were performed for X-ray tube voltages between 50 and 120 kVp.
RESULTS: Gd2O2S:Tb phosphor exhibited higher brightness and sharpness, as compared to the other phosphor materials, for all screens and X-ray tube voltages used. Best Gd2O2S:Tb performance was observed for thin screens and high tube voltages. La2O2S:Tb exhibited higher BSI values than Y2O2S:Tb for medium and high tube voltages.
CONCLUSION: Results showed that phosphor materials of high X-ray detection and X-ray-to-light conversion properties exhibit high BSI values indicating that BSI may provide a means of phosphor performance evaluation for imaging applications.

DOI10.3109/02841859909177741
Alternate JournalActa Radiol
PubMed ID10080737